40
Converting Input Values (Scaling Function)
When using
[2-Point]
: Specify two input values and those after scaling.
Enter two voltage values of an inputted signal in the
[Input1]
and
[Input2]
boxes.
Enter each of two scaled value in the
[Scale1]
and
[Scale2]
boxes.
−9.9999E+29
to
9.9999E+29
The instrument converts (scales) measured values in
volts into values in the specified measurement unit.
Example:
2 points of values
in volts
Values after scaled
V
H
: Higher potential
point
A
H
: Value for the higher
potential point
V
L
: Lower potential
point
A
L
: Value for the lower potential
point
Unit: A
[A]
A
H
A
L
V
L
V
H
Scaling values using a
conversion ratio and offset value,
both of which are calculated
using two points
Values after scaled
Actual measured values
[V]
When changing the value with the
[Ratio]
setting, the instrument does not change the values
for V
L
or V
H
, but the values for A
L
and A
H
, all of which have been specified with
[Method]
set
to
[2-Point]
.
When using
[Sensor]
•
Click the
[Sensor]
box, and then choose a model name of a current sensor or differential probe from
the list.
•
Click the
[Range]
box, and then choose a measurement range from the list.
Sensor
Range
Model 3273-50
30 A
Model 3274
150 A
Model 3275
500 A
Model 3276
30 A
Model 3283
10 mA
, 100 mA, 1 A, 10 A, 200 A
Model 3284
20 A
, 200 A
Model 3285
200 A
, 2000 A
Model 9010-50
10 A
, 20 A, 50 A, 100 A, 200 A, 500 A
Model 9018-50
10 A
, 20 A, 50 A, 100 A, 200 A, 500 A
Model 9132-50
20 A
, 50 A, 100 A, 200 A, 500 A, 1000 A
Model 9322
–
Model 9657-10
10 A
Model 9675
10 A
CT6700, CT6701
5 A
CT6710, CT6711
500 mA, 5 A, 30 A
Setting example:
To display values measured with Model 9018-50 Clamp on Probe using the 10 A range as values in
amperes (A)
Sensor
:
Model 9018-50
Range
:
10 A
Unit
A
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......