63
Configuring Measuring-Module-Specific Settings
4
Click the
[A.A.F.]
box, and then choose a anti-aliasing filter setting from a list.
The anti-aliasing filter can prevent aliasing distortion that may be produced during FFT calculation. The cutoff
frequency automatically changes according to the sampling rates or frequency range (for the FFT function)
settings.
Off
Disables the anti-aliasing filter.
On
Enables the anti-aliasing filter.
(Disabled when the external sampling is used, or the sampling rate is set at 100 kS/s
or faster)
5
Click
[Sensitivity]
box, and then enter sensor sensitivity.
You can enter sensor sensitivity to two decimal places. For a charge-output acceleration sensor or non-TEDS-
compliant sensor, enter its sensitivity marked on the sensor, which represents sensitivity per meter per second
squared.
6
Click
[Close]
.
The setting dialog box closes.
Setting example for sensor sensitivity
Example 1
For a sensor with its sensor sensitivity per meter per second squared marked
Sensor sensitivity
Setting value
1.08 pC/(m/s
2
)
1.08
Example 2
For a sensor with its sensor sensitivity per gravity (G) marked.
For a sensor with its sensitivity per gravity (G) marked, enter a quotient of the marked sensitivity
divided by 9.8 m/s
2
.
Sensor sensitivity
Setting value
For sensor sensitivity of 64 pC/G
64.0 / 9.8 = 6.53061... pC/(m/s
2
)
6.531 (to two decimal places)
To convert a unit from meter per second squared into gravity (G)
The instrument measures charge quantities per meter per second squared. You can convert such
charge quantities into those per gravity (G) using the scaling function.
Refer to “3.2 Converting Input Values (Scaling Function)” (p. 38).
Configure the scaling setting as follows;
Example 1
Specifying a conversion ratio
Ratio
0.1020E+00 (= 1/9.8)
Offset
0.0000E+00
Units
G
Example 2
Specifying two points
Input1
9.8000E+00
Scale1
1.0000E+00
Input2
0.0000E+00
Scale2
0.0000E+00
Units
G
3
Advanced Functions
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......