![Hioki MR8740-50 Скачать руководство пользователя страница 152](http://html1.mh-extra.com/html/hioki/mr8740-50/mr8740-50_instruction-manual_2137326152.webp)
146
Configuring the Numerical Calculation Settings
Calculation type
Setting
Description
Example of screen
Arithmetic
operations
Calculation number 1
,
Calculation number 2
(No. 1 through No. 107)
Allows you to specify
two numerical
calculation numbers to
use for calculations.
Target channel
Allows you to specify a
target channel that has
a numerical calculation
number to use for
calculations.
Operator
(
+
,
−
,
×
,
÷
)
Allows you to choose
among the four basic
arithmetic operations.
*: The range you can specify are between −9.9999E+29 and −1.0000E-29, 0, and b1.0000E-29 and
+9.9999E+29. A value to five significant figures can be set (a value to 10 significant figures can be set for a
time of the level at time setting).
•
With the
[Period]
,
[Frequency]
,
[Rise time]
, and
[Fall time]
calculations, the instrument may not
be able to perform calculations depending on the condition of waveform data.
•
With the
[Period]
and
[Frequency]
calculation, correct measurement results may not be obtained
if the filter setting is close to half the period (an integer multiple of an actual period may be
calculated).
•
When the scaling is enabled, the instrument scales waveform data before calculation. Units for
parameter values are those set in the scaling function.
•
If a waveform across a calculation target channel exceeds the measurable range (overrange
waveform), the instrument substitutes the upper or lower measurement limit for values that
exceed the measurable range before calculation.
Refer to “3.2 Converting Input Values (Scaling Function)” (p. 38).
Settings: Calculations for which
[Stat.]
can be set
Period, Frequency, Rise time, Fall time, Pulse width, Duty ratio, Time difference, Phase contrast, +Width,
−
Width, and Burst width
Top
Makes calculations within the calculation range in the initial condition.
Average
Calculates the average value of the calculation results for the data within the
calculation range.
Max
Calculates the maximum value of calculation results for data within the calculation
range.
Min
Calculates the minimum value of calculation results for data within the calculation
range.
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......