Fine-Adjusting Input Values (Vernier Function)
3.3
Fine-Adjusting Input Values (Vernier Function)
You can freely fine-adjust input voltage on the waveform screen. When recording physical values,
such as noise, temperature, and acceleration, with sensors, you can adjust those amplitudes, which
facilitates calibration.
Normal display
Waveform processed by the vernier function
When an input voltage of 1.2 V is displayed
as a voltage of 1.0 V
1.0 V
1.2 V
> [Channel]
1
Click
[Vernier]
.
The adjustment keypad appears.
2
While observing the waveform, you can fine-adjust its amplitude by clicking
[−−]
,
[−]
,
[+]
,
and
[++]
.
Click
[C]
to revert the fine-adjusted amplitude to its original.
•
The adjustable range is from 50% to 200% of an original waveform.
•
You cannot check if waveforms are adjusted by the vernier function by observing waveforms only.
•
Waveform data (data saved as files) is that adjusted by the vernier function.
3
Advanced Functions
Содержание MR8740-50
Страница 2: ......
Страница 6: ...Contents iv...
Страница 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Страница 86: ...80 Configuring Generator Module Specific Settings...
Страница 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Страница 164: ...158 Numerical Calculation Types and Descriptions...
Страница 182: ...176 Displaying Waveforms...
Страница 186: ...180...
Страница 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Страница 224: ...218 External Sampling EXT SMPL...
Страница 232: ...226 Information for Reference Purposes...
Страница 236: ...230 Index...
Страница 237: ......
Страница 238: ......