4.3 Setting DC Resistance Measurement
69
4
Chapte
r 4
LCR Function
There are the following two methods for setting the measurement range.
4.3.2
Setting the Measurement Range
1
Setting AUTO Ranging
AUTO
The most suitable test range is set automatically.
(This allows the most suitable measurement range to be set when, for example,
measuring a test sample whose nature is unknown.)
HOLD
The test range is fixed, and may only be altered manually.
(When the range is fixed, high-speed measurement is possible.)
Procedure
LCR Initial Screen
Press
.
2
1
Rdc Basic Settings
Rdc Basic Settings
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......