7.3 Compensating Values to Match Reference Values (Load Compensation)
296
Setting the DC Bias
If you make a mistake during input:
press
to cancel the input and start again.
DC Bias Setting
2.
Select ON/ OFF for the DC bias.
Sets the DC bias to OFF.
Sets the DC bias to ON.
3.
Use
or
to enter the DC bias value.
Settable range: 0.00 to 2.50 V
4.
Press
to close the setting screen.
DC Bias Setting
1.
Press
.
• If the compensation frequency, compensation range, and compensation level are not set, the
DC bias setting cannot be set.
• When DC is selected for the compensation frequency setting, the DC bias setting cannot be
set.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......