4.3 Setting DC Resistance Measurement
67
4
Chapte
r 4
LCR Function
When the measurement value is outside the guaranteed accuracy range, the following icon
appears at the top of the screen.
In this case, you should consider the following possible causes, and you should either
change the test conditions while checking the accuracy assured ranges "13.2 Measure-
ment Range and Accuracy" (p. 395), or you should consider the measured values as values
for reference.
• Perhaps the test signal level is too low, increase the test signal level.
• If the current measurement range (during HOLD setting) is not appropriate, set again in the AUTO
range, or change the range by manual.
3
To select the measurement signal level.
The accuracy of testing varies according to the test
signal level.
See
"13.2 Measurement Range and Accuracy" (p. 395)
Measurement Signal Level Setting
Voltage level between test sample termi-
nals (p. 45)
Current level through test sample
(p. 47)
4
Use
or
to enter the voltage or current
value.
5
Press
to close the setting screen.
Measurement Signal Level Setting
See
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......