4.2 Setting Basic Settings of Measurement Conditions
45
4
Chapte
r 4
LCR Function
Relationship between the measurement signal level of the instrument and the sample is as follows.
• When normal mode (p. 54)
About the test signal level
For setting range and accuracy
Output impedance
Object under
test
CC
CV
H
L
V
Constant current level (CC)
You should select this if you wish
to set the current passing through
the object to be tested to a con-
stant value.
Constant voltage level (CV)
You should select this if you wish to set the
voltage across the terminals of the object
to be tested to a constant value.
Open circuit voltage level (V)
This voltage value is the value which is applied across the two terminals of
the series combination of the object which is being tested and the output
impedance. As for the voltage which is applied across the terminals of the
object which is being tested (by itself), if required, you should either check
the monitor voltage value, or select constant voltage (CV) and set a voltage
value across these terminals.
Open circuit voltage level (V) and Constant voltage level (CV) setting
Measurement frequency
setting range
Open circuit voltage
setting range
Open circuit voltage
accuracy
4 Hz to 1.000 MHz
0.005 V to 5.000 V
±10%±10 mV
1.0001 MHz to 5.000 MHz
0.010 V to 1.000 V
±20%±10 mV
0.010 V
0.005 V
5.000 V
1.000 V
4 Hz
5 MHz
1 MHz
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......