4.2 Setting Basic Settings of Measurement Conditions
54
This function operates when measurement is being performed in low impedance high accuracy mode (p.
53). It detects whether each of the L
POT
, H
POT
and L
CUR
terminals is disconnected.
If a terminal is not connected properly, an indication like the following is displayed.
Check the connections again.
See
"2.4 Connecting the Measurement Cables, Probes, or Fixture" (p. 30)
The error is also output via EXT I/O.
See
"Chapter 11 External Control" (p. 363)
With low Z high accuracy mode, the possible setting ranges of the frequency and measure-
ment signal level differ. Refer to the following.
Settable range of measurement signal level: (p. 45)
Low Z high accuracy mode is only valid when the frequency is between 4 Hz and 100 kHz.
Detecting Disconnection of 4 Terminals
(When Low Z High Accuracy Mode)
When an element other than a capacitor or a capacitor with a low DC resistance is measured
while using DC bias, the contact check function may not work properly.
No.
Measurement
range
to 1 kHz
to 10 kHz
to 100 kHz
to 1 MHz
to 5 MHz
1
100 M
2
10 M
3
1 M
4
100 k
5
30 k
6
10 k
7
3 k
8
1 k
9
300
10
10
11
1
12
100 m
Normal mode only
(setting not possible for low Z high accuracy mode).
None
Low Z high accuracy mode/ normal mode
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......