![Hioki IM3570 Скачать руководство пользователя страница 309](http://html1.mh-extra.com/html/hioki/im3570/im3570_instruction-manual_2137305309.webp)
7.4 Compensating Measurement Cable Errors (Cable Length Compensation)
301
7
Chap
ter 7 Erro
r Com
pensatio
n
With high frequency measurement, the influence of the cable results in large measurement errors.
Setting the cable length enables you to reduce the measurement errors.
Use a coaxial cable with 50
impedance.
7.4 Compensating Measurement Cable
Errors
(Cable Length Compensation)
• When a 2 m cable is used, the guaranteed accuracy is 1 MHz.
• When a 4 m cable is used, the guaranteed accuracy is up to 100 kHz.
LCR
ANALYZER
Procedure
Compensation Screen
1
2
Select the cable length to be used, and press
to confirm the selection.
3
Press
to close the setting screen.
LCR Initial Screen
Cable Length Compensation Setting
Select this when using a direct-coupled fix-
ture or the like.
Select this when the cable length is 1 m, 2m,
or 4 m.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......