2.4 Connecting the Measurement Cables, Probes, or Fixture
30
Be sure to read the “Usage Notes” (p. 7) before connecting measurement
cables, probes or test fixture.
Connect your measurement cables, optional Hioki probes or test fixture to the
measurement terminal
s.
Refer to
for details.
See the instructions provided with the fixture for operating details.
2.4 Connecting the Measurement
Cables, Probes, or Fixture
• Basically, when you make a probe yourself, it may not be able to satisfy the specifications
of this instrument.
See:
• If all four terminals are disconnected, a meaningless number may be displayed on the
unit.
• If all four terminals are disconnected, a measurement signal and DC voltage of 6 V may
be generated at the H
POT
terminal, and a DC voltage of 6 V at the L
POT
terminal.
Connecting a measurement cable/fixture
Connect directly to the measurement jacks with
the label side up, and affix with the levers on the
left and right.
Points to pay attention to when making your own probe
•
Use 50
coaxial cable for the measurement cable.
•
Ensure that the length of the cable is the same as that set for the instrument.
•
Make the portion of the core wire that is exposed as short as possible.
•
Connect the H
CUR
, L
CUR
, H
POT
, and L
POT
shield pairs at the measurement object side.
(Ensure that a shield is not connected to a core wire.)
Measurement Terminal
Configuration
Fixture
Normal mode
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......