5.3 Normal Sweep
147
5
Chapte
r 5
ANAL
YZER Functi
o
n
Fixing sweep parameter and setting measurement to be per-
formed at a set time interval (Time interval measurement)
If you make a mistake during input:
press
to cancel the input and start again.
1.
Press
and use the numeric keypad to set the
start value of the sweep.
Settable range: 4 Hz to 5 MHz
2.
Press
,
, or
key to confirm the setting.
3.
Press
and use the numeric keypad to set the
measurement time interval.
Settable range: 0 s to 10000 s
4.
Press
key to confirm the setting.
5.
Press
and use the numeric keypad to enter the
number of measurements.
Settable range: 2 to 801
6.
Press
key to confirm the setting.
7.
Press
key to confirm the setting.
INTERVAL Setting
When the sweep range setting method is changed, the measurement interval for INTERVAL mea-
surement is reflected in the point delay time.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......