3.1 When LCR Measurement Mode
33
3
Chapter 3
Meas
ure
ment Exa
mple
The following shows measurement examples of the LCR measurement mode and analyzer measurement mode.
Necessary items: 9263 SMD test fixture
Laminated ceramic capacity you want to measure
Measurement
Example
Chapter 3
3.1 When LCR Measurement Mode
LCR
ANALYZER
Measuring a Laminated Ceramic Capacitor
1
Connect the 9263 SMD test fixture to the measurement terminals.
For the connection procedure, refer to the
instruction manual supplied with the fixture.
2
Set the measurement conditions.
Select the item you want to set, and set it as
follows.
LCR Basic Settings
Measurement frequency: 1 kHz(p. 41)
Measurement signal level: 1 V (p. 43)
Measurement range: AUTO (p. 49)
Measurement speed: MED(p. 59)
Voltage and current limit: (p. 60)
Average: OFF (p. 62)
Trigger delay: 0 s (p. 64)
Trigger: INT(p. 55)
DC bias: OFF (p. 57)
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......