4.2 Setting Basic Settings of Measurement Conditions
60
Depending on the measurement signal level, in some cases it is possible to damage the sample which is
being tested by applying to it a voltage or a current greater than its rated value.
For this reason, set a limit value to restrict the voltage that can be applied to the test sample or current that
can flow to the test sample.
4.2.7 Setting the Voltage/Current Limit
When open circuit voltage (V) or
constant voltage (CV) is set:
Set the current limit.
When constant current (CC) is set:
Set the voltage limit.
Procedure
LCR Initial Screen
• The present situation can be checked using the monitor display.
• The monitor display is different for V, CV, and CC.
First set the measurement signal level, and thereafter set the voltage or current limit.
The setting for voltage or current limit changes automatically to current or voltage limit,
according to the present measurement signal level setting.
See
"4.2.2 Setting the Measurement signal level" (p. 43)
1
LCR Basic Settings
3
When the measurement signal level is a
voltage (V, CV)
Current Limit Setting
When the measurement signal level is a
current (CC)
Voltage Limit Setting
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......