5.3 Normal Sweep
166
For the point delay setting, set the delay time for each sweep point.
See
"5.2.5 Setting the Trigger Delay" (p. 140)
5.3.6 Setting the Point Delay
Procedure
Press
.
Analyzer Initial Screen
Analyzer Basic Settings
1
Analyzer Basic Settings
2
Use
or
to enter the delay time.
4
3
Press
to close the setting screen.
Settable range: 0.0000 s to 10000 s with resolution
of 0.1 ms
Point Delay Setting
When you want to cancel the point delay
function:
Press .
The setting value is cleared to 0.