13.1 General Specifications
388
Measurement signal level
(1) Open circuit terminal voltage (V) mode and constant voltage (CV) mode
• Level range
Normal mode: 5 mV to 5 V, maximum 50 mA (up to 1.0000 MHz)
10 mV to 1 V, maximum 10 mA (from 1.0001 MHz)
Low Z high accuracy mode: 5 mV to 1 V, maximum 100 mA (up to 100.00 kHz)
• Setting resolution 1 mV steps
• Setting accuracy
±
10%
of setting
±
10 mV (up to 1.0000 MHz)
±
20%
of setting
±
10 mV (from 1.0001 MHz)
(2) Constant current (CC) mode
• Level range
Normal mode: 10
A to 50 mA, maximum 5 V (up to 11.0000 MHz)
10
A to 10 mA, maximum 1 V (from 1.0001 MHz)
Low Z high accuracy mode: 10
A to 100 mA, maximum 1 V (100 m
and 1
ranges
of 100 kHz or less)
10
A to 10 mA (other than the above)
• Setting resolution 10
A steps
• Setting accuracy
±
10%
of setting
±
10
A (up to 1.0000 MHz)
±
20%
of setting
±
10
A (from 1.0001 MHz)
Measurement range
The measurement range is determined according to impedance Z.
The values of the other measurement items can be calculated.
Ranges: 100 m
, 1
, 10
, 300
, 1 k
, 3 k
, 10 k
, 30 k
, 100 k
, 1 M
,
10 M
, 100 M
(12 ranges)
• The guaranteed accuracy range differs depending on the measurement conditions.
• Out of guaranteed accuracy is displayed when out of the ranging range
OVERFLOW or UNDERFLOW is displayed when out of the A/D input range
1.Basic Specifications
Measurement range
Guaranteed Accuracy Range
AUTO Ranging Range
100 M
8 M
to 200 M
8 M
to 999.999 M
10 M
800 k
to 100 M
800 k
to 10 M
1 M
80 k
to 10 M
80 k
to 1 M
100 k
24 k
to 1 M
24 k
to 100 k
30 k
8 k
to 300 k
8 k
to 30 k
10 k
2.4 k
to 100 k
2.4 k
to 10 k
3 k
800
to 30 k
800
to 3 k
1 k
240
to 10 k
240
to 1 k
300
8
to 300
8
to 300
10
800 m
to 10
800 m
to 10
1
80 m
to 1
80 m
to 1
100 m
1 m
to 100 m
0
to 100 m
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......