7.3 Compensating Values to Match Reference Values (Load Compensation)
295
7
Chap
ter 7 Erro
r Com
pensatio
n
Setting the Type and Value of the Compensation Level
2.
Select the compensation level type.
Open voltage level (p. 45)
Voltage level between test sample terminals (p. 45)
Current level between test sample terminals (p. 47)
Level Setting
3.
Use
or
to enter the voltage or current value.
For the possible compensation setting ranges, see the
following figures.
4.
Press
to close the setting screen.
Level Setting
1.
Press
.
V, CV
Frequency
LOW Z
Range
V, CV
4.00 Hz to
1.0000 MHz
OFF
Entire
range
0.005 to
5.000 V
1.0001 MHz to
5.0000 MHz
OFF
Entire
range
0.010 to
1.000 V
4.00 Hz to
1.0000 MHz
ON
Entire
range
0.005 to
1.000 V
1.0001 MHz to
5.0000 MHz
ON
Entire
range
0.010 to
1.000 V
AC Load Compensation
CC
Frequency
LOW Z
Range
CC
4.00 Hz to
1.0000 MHz
OFF
Entire range
0.01 m to
50.00 mA
1.0001 MHz to
5.0000 MHz
OFF
Entire range
0.01 m to
10.00 mA
4.00 Hz to
100.00 kHz
ON
100 m
, 1
0.01 m to
100.00 mA
Range of
10
or more.
0.01 m to
10.00 mA
100.01 kHz to
5.0000 MHz
ON
Entire range
0.01 m to
10.00 mA
V, CV
LOW Z
Range
V, CV
OFF
Entire range
0.1 V to 2.50 V
ON
Entire range
0.1 V to 1.00 V
DC Load Compensation
CC
LOW Z
Range
CC
OFF
Entire range
0.01 m to 25.00 mA
ON
100 m
, 1
0.01 m to 100.00 mA
Range of 10
or more.
0.01 m to 25.00 mA
If the compensation range is not set, the compensation level cannot be set.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......