5.10 Equivalent Circuit Analysis Function
238
The optional IM9000 Equivalent Circuit Analysis Firmware is required in order to use the equivalent circuit
analysis function. You can verify whether the IM9000 is installed on the Version Confirmation screen.
See
"8.2 Checking the Version of the Instrument" (p. 306)
The equivalent circuit analysis function estimates equivalent circuit constants based on measurement results.
The IM3570 can estimate 3-element model and 4-element model constants as illustrated below. By using the
simulation function, you can display frequency characteristic ideal values using estimation results or user-con-
figured constants.
By using the comparator function, you can judge whether estimation results fall inside the judgment area.
*Typical frequency characteristics graphs
For models A through D, the horizontal axis is logarithmic, the vertical axis (Z) is logarithmic, and
is linear.
For model E, the horizontal axis is linear or logarithmic, the vertical axis (Z) is logarithmic, and
is linear.
5.10 Equivalent Circuit Analysis Function
5.10.1 About the Equivalent Circuit Analysis Function
Equivalent circuit model
Representative frequency
characteristics*
Example sample
A
Inductor: Inductor
with
high
core loss and low
ESR
B
Inductor: Inductor
with
compar-
atively high ESR
Resistor: Resistor
with
low
resistance value and
significant wiring
inductance effect
C
Capacitors: Capacitor with signifi-
cant leak resistance
effect
Resistor: Resistor
with
high
resistance value and
significant stray
capacitance effect
D
Capacitors: Typical capacitor
E
Piezoelectric element
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......