14.3 Error display
407
14
Chapter 14
Mainte
nance
and Service
When any of the following indications appear on the screen, check the corresponding reference
page.
14.3 Error display
Error display
Description
Remedy and Reference
The internal temperature of the instru-
ment is increasing.
Turn off the power and check whether
the ventilation holes are blocked. (p. 11)
The sum of the AC measurement signal
and DC measurement signal exceeds
the upper limit.
Reduce the AC signal level or DC signal
level. (p. 43)
If the sum of the AC signal level and DC
signal level exceeds the upper limit
when the signal level mode is changed,
the AC signal level is forcefully con-
trolled so that it becomes within the set-
ting range.
Reduce the AC signal level or DC signal
level. (p. 43)
The life of the RAM backup battery has
ended.
The instrument needs to be repaired.
Contact your dealer or Hioki representa-
tive.
This is displayed when a measurement
value is outside of the guaranteed accu-
racy range.
Increase the measurement signal level
or change the measurement range to
one that matches the impedance of the
element to be measured.(p. 43), (p. 49)
• This is displayed when load compen-
sation is enabled and the load com-
pensation frequency does not match
the current measurement frequency.
• This is displayed when constant volt-
age measurement and constant cur-
rent measurement cannot be
performed.
• When load compensation: Match the
current measurement frequency to
the compensation frequency.(p. 288)
• When constant voltage measure-
ment or constant current measure-
ment: Reduce the constant voltage
level or constant current level. (p. 45)
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......