Appendix
Appendix 7 Series Equivalent Circuit Mode and Parallel Equivalent Circuit Mode
A
11
The instrument measures the current flowing to the test sample and the voltage at both ends of the test sam-
ple, and determines Z and
. Other measurement items such as L, C, and R are calculated from Z and
. At
this time, the mode for calculation becomes series equivalent circuit mode if the resistance components for C
(or L) are assumed to be in series, and the mode becomes parallel equivalent circuit mode if the resistance
components for C (or L) are assumed to be in parallel. It is, therefore, necessary to select the correct equiva-
lent circuit mode to reduce errors because the calculation expression differs for series equivalent circuit mode
and parallel equivalent circuit mode.
Generally, for measurement of a low impedance device (approx. less than 100
) like a large capacitance
capacitor or a low inductance, a seriesequivalent circuit mode will be selected. While, for a high impedance
device (approx. more than 10 k
) like a small capacitance capacitor or a high inductance, a parallel-equiva-
lent circuit mode will be selected. When you are not sure about selection of circuit mode, please ask the parts
maker. (ex. a impedance approx. between 100
and 10 k
)
Appendix 7
Series Equivalent Circuit Mode and
Parallel Equivalent Circuit Mode
Series equivalent circuit
Parallel equivalent circuit
C (or L)
C (or L)
Because measurement value in each equivalent circuit mode is obtained through calcula-
tion, measurement values of both modes can be displayed. However, please note that the
appropriate equivalent circuit depends on the test sample.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......