12.1 Connecting the Printer
381
12
Cha
p
ter 12 Printing
Connecting the Printer
Recommended printer
The IM3570 has been verified to operate properly with Sanei Electric Inc.'s model SD1-31S thermal printer.
The requirements for a printer to be connected to the instrument are as follows.
Confirm compatibility and make the appropriate settings on the printer before connecting it to the instrument.
See
"12.2 Instrument and Printer Settings" (p. 383)
• Interface ....................... RS-232C
• Characters per line....... At least 45
• Communication speed . Initial value: 9,600 bps
• Data bits ....................... 8bit
• Parity............................ none
• Stop bits ....................... 1bit
• Flow control.................. none
Printing
Chapter 12
• Measurement values and
comparator decisions
• Screens
Connecting the printer
to the instrument
Make instrument
settings (p. 383)
Make printer
settings
12.1 Connecting the Printer
Because electric shock and instrument damage hazards are present, always follow
the steps below when connecting the printer.
• Always turn off the instrument and the printer before connecting.
• A serious hazard can occur if a wire becomes dislocated and contacts another con-
ductor during operation. Make certain connections are secure.
•
As much as possible, avoid printing in hot and humid environments. Otherwise, printer
life may be severely shortened.
•
Use only compatible recording paper in the printer. Using non-specified paper may not
only result in faulty printing, but printing may become impossible.
•
If the recording paper is skewed on the roller, paper jams may result.
The communication speed and flow control can be changed with instrument settings. However,
the IM3570 and printer must be configured with the same settings.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......