4.5 Setting Application Settings
108
This function is for outputting a measurement terminal connector error when the measurement result is high
relative to the set judgment reference value. The setting value can be set as an absolute value, and the error
is output via the EXT I/O.
See
"Chapter 11 External Control" (p. 363)
The judgment reference is calculated from the nominal value (range name) of the current measurement range
and the judgment reference value as shown below.
Judgment reference = Nominal value of current measurement range
×
Judgment reference value (%)
4.5.3 Detecting OPEN during 2-terminal Measurement
(HIGH-Z Reject Function)
Example Current measurement range: 30 k
Judgment reference value: 150%
Judgment reference = 30 k × 1.50 = 45 k
Procedure
LCR Initial Screen
Press
.
2
Application Settings
Application Settings
1
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......