5.2 Setting Basic Settings of Measurement
140
Set the delay time from when a trigger is input until measurement starts.
There are two delay settings: “Trigger Delay” and “Point Delay.”
With this item, only the setting for the trigger delay is configured.
5.2.5 Setting the Trigger Delay
When sequential sweep
When repeat sweep
When step sweep
External trigger
End of measurement
End of measurement
End of measurement
End of measurement
End of measurement
Internal trigger
Return to measurement of sweep point 1
External trigger
External trigger
External trigger
External trigger
Trigger delay
Point delay
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
last sweep point
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
last sweep point
Measurement of
sweep point 1
Measurement of
sweep point 2
Measurement of
sweep point 3
Measurement of
last sweep point
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......