7.3 Compensating Values to Match Reference Values (Load Compensation)
289
7
Chap
ter 7 Erro
r Com
pensatio
n
• When all of the compensation conditions match for load compensation, compensation is
performed under the optimal conditions.
• If the current measurement frequency and compensation frequency do not match, an
error like the following is displayed in the initial screen.
• If a condition other than the compensation frequency does not match, compensation is
performed but a warning like the following is displayed in the initial screen.
• When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and
processed by the OPEN or SHORT compensation.
• In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
• If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......