![Hioki IM3570 Скачать руководство пользователя страница 375](http://html1.mh-extra.com/html/hioki/im3570/im3570_instruction-manual_2137305375.webp)
11.1 External Input/Output Connector and Signals
367
11
Chapte
r
1
1 External
Co
ntro
l
You can select rising or falling for the valid edge of a trigger.
See
Signal Descriptions
Input
TRIG
• When the trigger setting is the external trigger
, measurement is performed
once with the falling (ON) or rising (OFF) edge of the TRIG signal. The edge direction
can be set in the setting screen. (Initial value: Falling (ON))
See:
"Enabling Trigger Input for during Measurement and Setting the Valid Edge of Trig-
ger Input" (p. 117), (p. 228)
• When the trigger source is set to the internal trigger
, trigger measurement is
not performed.
• You can set whether to enable or disable TRIG signal input during measurement (during
output of the EOM signal (HI)).
See:
"Enabling Trigger Input for during Measurement and Setting the Valid Edge of Trig-
ger Input" (p. 117), (p. 228)
LD0 to LD6
Selects the number of the panel to load.
If a trigger signal is input in external trigger mode, the selected panel is loaded and used for
measurement.
LD-VALID
Inputs a negative logic signal from an external device so that the selected panel number is
recognized as valid.
After TRIG input, maintain a Low level until INDEX is outputted.
CALIB
When the DC adjustment function for during DC resistance measurement is set to OFF, the
offset value generated in the internal circuit can be obtained at the desired timing.
After TRIG input, maintain a Low level until INDEX is outputted.
PIN No.
LD6
LD5
LD4
LD3
LD2
LD1
LD0
Panel 1
0
0
0
0
0
0
1
Panel 2
0
0
0
0
0
1
0
Panel 4
0
0
0
0
1
0
0
Panel 8
0
0
0
1
0
0
0
Panel 16
0
0
1
0
0
0
0
Panel 32
0
1
0
0
0
0
0
Panel 64
1
0
0
0
0
0
0
Panel 127
1
1
1
1
1
1
1
Panel128
0
0
0
0
0
0
0
0: (HIGH: 5 V to 24 V), 1: (LOW: 0 V to 0.9 V)
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......