7.1 Setting Open Circuit Compensation
279
7
Chap
ter 7 Erro
r Com
pensatio
n
If compensation fails, a window such as the following will be displayed.
When an error message appears and compensation has stopped (when
is touched), open cir-
cuit compensation is turned OFF.
The open circuit compensation process is quite sensitive to noise - both noise originating externally and
induced noise.Therefore, if open circuit compensation has been interrupted with a fault, you should check
the following points before starting the compensation process again:
• Check that the test cables are properly connected.
• Check that nothing is connected to the test cables. (Open circuit compensation cannot be performed
while any test sample is connected to the test cables.)
• Check that the test leads are arranged as closely as possible to their configuration in which measure-
ment will be performed.
• During the compensation process, be sure not to disturb the test cables or to move your hand near
them.
• Execute the guarding process.
See
"Appendix 2 Measurement of High Impedance Components"(p. A3)
Press
in Step
3
of [Open Compensation Setting] (p. 274) to make the compensation data for up
until now invalid and return to the [Initial screen].
When Open Compensation Failed
When You Want to Make Open Circuit Compensation Data Invalid
Error
Open Compensation Setting
The compensation values that are stored internally are not cleared by the operation
described above. When ALL or SPOT is selected, the stored compensation values can be
used.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......