7.1 Setting Open Circuit Compensation
278
The compensation values from last time are dis-
played in a confirmation screen.
(If compensation has never been performed, the
compensation values become 0.)
5
Spot Compensation Value Setting
6
Compensation starts.
Compensation value acquisition time:
Approximately 5 seconds
Performing Open Compensation
Check that the measurement cable is in an open circuit
state.
Press .
When you want to cancel compensation:
Press .
Compensation is cancelled and the compensation
screen is redisplayed.
(The open circuit compensation value state from last
time remains.)
When you do not want to acquire the compen-
sation values:
Press .
The setting screen is redisplayed, and the compensation
values from last time become valid.
Spot Compensation
7
When compensation failed:
• If compensation ends normally, the conductance
and susceptance are displayed.
• Compensation can be performed for impedances of at
least 1 k
.
8
Press
to close the setting screen.
Indicates the
compensation
numbers.
You can use
and
to check the Conduc-
tance, susceptance of each compensation point.
Indicates the
measurement
frequencies.
Indicates the compensation results.
(Conductance, susceptance)
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......