4.5 Setting Application Settings
111
4
Chapte
r 4
LCR Function
The number of waves that can be monitored differs depending on the frequency set on the instrument.
×
:The monitoring the detection level function is not activated.
:
The monitoring the detection level function is activated for all acquired waves
.
1/2
:The monitoring function is activated for the first half of acquired waves.
1/4
:The monitoring function is activated for the first quarter of acquired waves.
Monitoring the Detection Level Function Operation by Set Frequency
3
Detection Level Monitoring Function Setting
Select ON/OFF for the function to monitor the
detection level.
Sets the detection level monitoring function to
OFF.
Sets the detection level monitoring function to
ON.
Use
or
to enter the limit value.
5
4
Press
to close the setting screen.
Settable range: 0.01 to 100.00%
Detection Level Monitoring Function Setting
Frequency
FAST
MED
SLOW
SLOW2
DC
×
4.00 Hz to 19.99 Hz
×
20.00 Hz to 100.00 Hz
×
100.01 Hz to 999.99 Hz
×
1.0000 kHz to 3.0000 kHz
×
3.0001 kHz to 10.000 kHz
×
10.001 kHz to 30.000 kHz
30.001 kHz to 100.00 kHz
100.01 kHz to 300.00 kHz
1/2
1/2
1/2
1/2
300.01 kHz to 1.0000 MHz
1/4
1/4
1/4
1/4
1.0001 MHz to 5.0000 MHz
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......