7.3 Compensating Values to Match Reference Values (Load Compensation)
288
Compensate measurement values to match the element that will be the reference.
With load compensation it is possible to calculate the compensation coefficient by measuring a reference
sample with known data and perform the compensation for the test data obtained from the target sample.
The compensation coefficient can be acquired using up to five compensation conditions. The reference value
of each compensation condition can be set independently.
The following seven compensation conditions should be set for each measurement frequency:
The compensation coefficient is computed from the reference values of Z and
obtained from the set values
and the actual data acquired from the reference sample at each of the compensation frequencies.
The measured values of Z and
are first compensated using the following equations, and then individual
parameters from the compensated Z and Éý values are employed.
7.3 Compensating Values to Match Refer-
ence Values (Load Compensation)
LCR
ANALYZER
Set the Z/ Cs/ Cp/ Ls/ Lp/ Rs reference
value selected for the parameter type.
Compensation
frequency
Define the measurement frequency
used to measure and compensate the
reference sample.
Compensation
Range
Set the range to compensate.
Compensation
Signal Level
Set the type and value of the signal level
to compensate.
DC Bias
Enable or disable DC bias and set the
value.
Parameter Type
Set the parameter to use as the ref-
erence value.
Reference
Value 1
Reference
Value 2
Set the
/ D/ Rs/ Rp/ Q/ X reference
value selected for the parameter type.
Acquires the current
measurement conditions.
Deletes the
compensation
conditions.
Compensation coefficient of Z = (Reference value of Z) / (Actual data of Z)
Compensation value of
= (Reference value of
) - (Actual data of
)
Z = (Z before compensation) × (Compensation coefficient of Z)
= (
before compensation) + (Compensation value of
)
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......