5.7 Judging Measurement Results (Comparator Function)
213
5
Chapte
r 5
ANAL
YZER Functi
o
n
14
Use the numeric keypad to set the left, right, upper, and lower limit values.
1.
Press
and use the numeric keypad to enter the
left limit value.
The possible setting range differs depending on the
sweep parameter.
Refer to the following for each of the parameters.
• Frequency: (p. 41)
• Open circuit voltage level: (p. 43)
• Voltage level between test sample terminals:
• Current level between test sample terminals:
2.
Press a unit key to confirm the setting.
3.
Press
and use the numeric keypad to enter the
right limit value.
The possible setting range differs depending on the
sweep parameter.
Refer to the following for each of the parameters.
• Frequency: (p. 41)
• Open circuit voltage level: (p. 43)
• Voltage level between test sample terminals:
• Current level between test sample terminals:
4.
Press a unit key to confirm the setting.
Left/Right Limit Value Settings
When you do not want to set the left, right,
upper, and lower limit values:
Press
.
If you make a mistake during input:
press
to cancel the input and start again.
Upper/Lower Limit Value Settings
5.
Press
and use the numeric keypad to set
the upper limit value.
Settable range: -9.999999G to 9.999999G
6.
Press
to confirm the setting.
7.
Press
and use the numeric keypad to set
the lower limit value.
Settable range: -9.999999G to 9.999999G
8.
Press
to confirm the setting.
9.
Press
to confirm the setting.
When you want to cancel the setting:
Press .
When you do not want to set a judgment
are
a: Press
.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......