4.4 Judging Measurement Results
87
4
Chapte
r 4
LCR Function
• The comparator judgment is made in the following order.
1.
If the measurement value is "OVER FLOW",
HI
is displayed.
(However, LO is displayed when the parameters are Y, Cs, Cp, G, and B.)
If the measurement value is "UNDER FLOW",
LO
is displayed.
(However, HI is displayed when the parameters are Y, Cs, Cp, G, and B.)
If the measurement value is “L NO CNTCT” or “H NO CNTCT”,
HI
is displayed.
2.
Whether the measurement value is higher than the lower limit value is judged, and
LO
is displayed if the judgment is NG.
3.
Whether the measurement value is lower than the upper limit value is judged, and
HI
is
displayed if the judgment is NG.
4.
When both 2 and 3 are satisfied,
IN
is displayed.
No test is performed to ensure that the upper limit value is greater than the lower
limit value, so no error message will be displayed if you set the upper limit value
and lower limit value the wrong way around.
• If the power is turned off while the comparator measurement screen is displayed, the
comparator measurement screen will be displayed when the instrument starts the next
time you turn the power on.
*
2: The following equation is used to calculate the
% value.
% =
|reference value|
measurement
×100
*
1: The following equation is used to calculate the comparison upper limit value and comparison lower limit value.
(In the case of the comparison lower limit value, if a value that is lower than the reference value is set, the minus
(-) sign is required for the percentage setting value.)
Upper limit comparison value (Lower limit comparison value) = reference value+ |reference value| ×
100
Percentage set value
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......