7.2 Short Circuit Compensation
283
7
Chap
ter 7 Erro
r Com
pensatio
n
The compensation values from last time are dis-
played in a confirmation screen.
(If compensation has never been performed, the
compensation values become 0.)
5
Performing Short Compensation
4
All Compensation
Compensation starts.
Compensation value acquisition time:
Approximately 2 minutes
Check that the measurement cable is in a short-circuit
state.
Press .
When you want to cancel compensation:
Press .
Compensation is cancelled and the compensation
screen is redisplayed.
(The short circuit compensation values from last time
remain.)
When you do not want to acquire the compen-
sation values:
Press .
The setting screen is redisplayed, and the com-
pensation values from last time become valid.
Short Compensation
6
When compensation failed:
• If compensation ends normally, the effective resis-
tance and reactance are displayed.
• The possible compensation range is 1 k
or less for
impedance.
Use
or
to check the effective resistance
and reactance of each compensation point.
7
Press
to close the setting screen.
Indicates the
compensation
numbers.
Indicates the compensation results.
(Effective resistance, reactance)
Indicates the
measurement
frequencies.
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......