The EXT I/O connector on the rear of the instrument supports external control by providing output of the end-
of-measurement and comparator decision signals, and accepting input of measurement
trigger and panel load signals. All signals are isolated by optocouplers (inputs and outputs share a common
signal ground (ISO_COM signal ground).)
Confirm input and output ratings, understand the safety precautions for connecting a control system, and use
accordingly.
External Control
Chapter 11
Signal input/output
Connect the instrument's
EXT I/O connector to the sig-
nal output or input device.
Make instrument settings
11.1 External Input/Output Connector and
Signals
To avoid electric shock or damage to the equipment, always observe the following
precautions when connecting to the EXT I/O terminals.
• Always turn off the power to the instrument and to any devices to be connected
before making connections.
• During operation, a wire becoming dislocated and contacting another conductive
object can be serious hazard. Make sure that connections are secure and use
screws to secure the external connectors.
• Ensure that devices and systems to be connected to the EXT I/O terminals/ are
properly isolated.
To avoid damage to the instrument, observe the following cautions:
•
Do not apply voltage or current to the EXT I/O terminals that exceeds their ratings.
•
When driving relays, be sure to install diodes to absorb counter-electromotive force.
•
Be careful not to short-circuit ISO_5V to ISO_COM.
See
Содержание IM3570
Страница 1: ...IM3570 Instruction Manual IMPEDANCE ANALYZER EN Dec 2018 Revised edition 8 IM3570A981 08 18 12H ...
Страница 2: ......
Страница 8: ...Contents vi ...
Страница 16: ...Operating Precautions 8 ...
Страница 34: ...1 3 Screen Configuration and Operation 26 ...
Страница 138: ...4 5 Setting Application Settings 130 ...
Страница 156: ...5 3 Normal Sweep 148 When the setting is or Procedure Analyzer Initial Screen Press 2 1 Example When ...
Страница 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Страница 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Страница 336: ...9 4 Deleting a Panel 328 ...
Страница 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Страница 418: ...14 4 Discarding the Instrument 410 ...
Страница 444: ...Appendix 13 Device Compliance Statement A26 ...
Страница 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Страница 448: ...Index4 ...
Страница 455: ......
Страница 456: ......
Страница 457: ......