8-58
Section 8-3 - Internal Component Replacement Procedures
8-3-9
LCD Flex Cable Replacement Procedure
8-3-9-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Flex Cable replacement
procedure.
8-3-9-2
Time Required
15 minutes
8-3-9-3
Preparation
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-3-9-4
LCD Flex Cable Removal Procedure
1) Disconnect the AC cable and battery.
2) Remove the Control Panel and Keyboard as described in the
Control Panel and Keyboard Removal
3) Remove the LCD Display Frame as described in the
LCD Display Frame Removal Procedure on
4) Remove the LCD Display as described in the
LCD Display Removal Procedure on page 8 - 53
.
5) Using both hands, carefully lift the LCD Display upwards and remove it from the Rear Cover & Latch
Assembly, as shown in
NOTE:
Take care not to damage the LCD Flex ribbon cable when performing this step.
Flex Cable Replacement Procedure
Figure 8-68 Removing the LCD Display