GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
4-36
Section 4-3 - Functional Checks
4-3-4-3
Plug and Scan Test
18
Press
OK
to close the message and then press
Done
.
A progress bar and message are displayed asking you
to wait.
19
Press
Alt+E
and select
CD/DVD Writable (H:\)
from the
menu.
The message
Media Ejected
is displayed.
20
Remove the DVD from the drive.
21
Verify that all of the saved exams and images of the
Test Patient
exist in the database.
The exams and images that were exported and
imported for the
Test Patient
should appear in the
database.
22
Select the
Test Patient
exams and press
Delete
.
A message is displayed prompting the user to confirm
the delete command.
23
Press
OK
to confirm delete (must confirm three times).
The
Test Patient
is deleted from patient list.
Table 4-25 Plug and Scan Test - Recommended
Task
Expected Result(s)
1
Press the
On/Off
button on the control console for no
more than 3 seconds.
The system enters
Standby
mode.
2
Turn OFF the system’s circuit breaker.
The system remains in
Standby
mode, indicated by the
blinking
On/Off
button LED.
3
Turn ON the system’s circuit breaker.
Press the
On/Off
Button.
The system returns quickly to the
Power-ON
state -less
than 30 seconds.
4
Turn OFF the system’s circuit breaker. Turn ON the
system’s circuit breaker after entering to Standby mode.
The system enters
Standby
mode and stays in this
mode. The
On/Off
button LED is blinking.
5
Turn OFF the system’s circuit breaker.
Press the
On/Off
button.
The system enters the
Power-OFF
state. The
On/Off
button LED is OFF.
Table 4-24 DVD/CD Writer Test (cont’d) - Recommended