7-32
Section 7-4 - Performing Front End (FE) Diagnostics
5.) Trackball to the
Start
button and press
Set
.
The Full RFI Test commences. While the test proceeds, the name of the test currently in progress
(Full RFI Test) is displayed in the
Execute
field and the current
Loop #
and specific test are shown
below. The messages
Init Done
and
Please
Wait
are displayed beside the progress bar in the
Status
area:
As the testing sequence progresses, the progress bar will advance to reflect the test progress. The
LED on the RFI status indicator in the Data Flow map will be highlighted in the appropriate color to
indicate the current test status, as follows:
•
Green:
All recommended tests for these components have been completed and no problems
were identified.
•
Red:
Problems were identified in these components during the performance of the test.
•
Yellow:
The tests executed so far on these components have passed, but not all
recommended tests have been performed.
6.) At any stage, trackball to the
Pause
button and press
Set
, if required.
7.) To resume testing (from the point where you paused), trackball to the
Start
button and press
Set
.
When the Full RFI Test sequence is complete:
-
the
Diagnostic
Test
window displays the Finished message
-
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in the example in
8.) If necessary, use the scroll bars and/or arrows to scroll through the list of
Test
results.
Note:
For the Full RFI test, four tests are performed in the test sequence; these are listed in the
. For each currently-selected test (for example,
RFI
FE Cache Test
) the corresponding report is listed in
Reports
on the right.
For details of each report, refer to the following explanations:
-
View Test Log - RFI Cache Test on page 7 - 35
-
View Test Log - RFI Demodulator Short Test on page 7 - 41
-
View Test Log - RFI Analog Circuits Test on page 7 - 38
-
View Test Log - RFI Audio Test on page 7 - 44
9) When done, trackball to the
Exit
button and press
Set
to close the View Report window.
Figure 7-12 View Test Log - Full RFI Test
Содержание Vivid in
Страница 2: ......
Страница 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii...
Страница 739: ......