Section 8-2 - External Component Replacement Procedures
8.) Press
Config
and log on as Adm - see
Log On to the System as ADM on page 4 - 10
.
9.) Select
About
, then click on the
HwVersion
tab.
10.)Verify that the correct ECG/Respiratory module Part # is displayed, as shown in
11.)Proceed to perform the following functionality tests:
-
Isolated Patient Lead (Source) Leakage – Lead-to-Ground on page 10 - 27
-
Isolated Patient Lead (Source) Leakage – Lead-to-Lead on page 10 - 28
-
Isolated Patient Lead (Sink) Leakage - Isolation Test on page 10 - 28
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the
Vivid
i
n/ Vivid q N system is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
-
ECG/Respiratory Test on page 4 - 25
Figure 8-22 Hardware Version Verification