Chapter 8 - Replacement Procedures
8-107
17.
Navigate to the
TCPIP
tab and press the
Set
button on
the control panel.
Connectivity dialog box - TCPIP tab activated:
18.
Navigate to the
Save Settings
button and press the
Set
button on the control panel.
A message is displayed prompting for system re-boot:
19.
Click
OK
and re-boot the system.
20.
If this is a new software version, an adhesive Rating
Plate is supplied with the software CD.
Adhere the plate in the appropriate position under the
Bottom Assembly.
21.
Proceed to perform the following functionality tests:
•
Software Configuration Checks on page 4 - 37
•
Calibration Tests on page 7 - 25
•
Full System Test on page 7 - 10
Table 8-6
Vivid
i
n/ Vivid q N Software Re-Installation (cont’d)
Step
Instruction
Expected Result