GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
4-26
Section 4-3 - Functional Checks
If applicable, proceed to run the
Respiratory Test
, as
described below.
8
Activate the Respiratory radio button.
The Respiratory signal is displayed in the Data area, as
shown in the example below:
9
Verify that a single clear Respiratory signal is displayed.
10
When done, trackball to the
Exit
button and press
Set
.
11
When prompted by the following message, select
Yes
or
No
as applicable.
12
Press the <
Esc
> button on the keyboard to close the
Diagnostic Test window.
13
When done, press
Exit
button
Table 4-12 ECG/Respiratory Test - Recommended
Step
Task
Expected Result(s)