Section 7-6 - Performing Back End Diagnostics on the System
10.)Check the Gain rotary as follows:
a.) When turning the rotary in a
clockwise
direction, each time you hear a click,
right
arrows are
displayed in the Test Screen.
b.) When turning in a
counter
-
clockwise
direction, each time you hear a click,
left
arrows are
displayed in the Test Screen.
Note:
When performing this step, make sure that a minimum of 32 clicks are heard when turning
the Gain rotary in both the
clockwise
and
counter
-
clockwise
directions to ensure that a
complete check is performed.
11.)Check the On/Off button.
a.) Press the button once; make sure the button is shown as depressed
in the Test Screen.
b.) Release the button; make sure the button is no longer showing as depressed in the Test
Screen.
12.)Check the Lights Control (Backlights and Active Lights), as follows:
a.) trackball to the Active Lights
button (under
Lights Control
) and press
Set
. Make sure that all
the corresponding LEDs are activated (illuminated green -
Patient
,
CW
,
Application
,
etc.
).
alternatively, check each one (one-by-one) individually.
b.) trackball to the Backlights
button (under
Lights Control
) and press
Set
. Make sure that all the
extended keyboard backlighting is correctly illuminated.
13.)Trackball to the Battery Light
button (top
right
of Test Screen) and press
Set
. Make sure that all the
corresponding LEDs is activated (illuminates green).
14.)Trackball to the Ext DC Light
button (top
right
of Test Screen) and press
Set
. Make sure that all the
corresponding LEDs is activated (illuminates green).
15) Perform the Trackball Test, as follows:
a.) Move the trackball; the position of the trackball cursor on the screen is reflected by a change
in the displayed X and Y coordinate values.
Note:
when the trackball cursor is positioned at the extreme top left of the screen:
X = 0 and Y = 0.
b.) As you move the trackball cursor, make sure the coordinate values
increase
/
decrease
in small
steps, without jumping more than two or three digits. If for example, the values increment by
steps of 10, this indicates that the trackball operation is faulty (possibly the trackball requires
cleaning, or replacement).
16) When done with all the above tests, trackball to the
OK
button and press
Set.