Section 7-7 - InSite ExC
7-7-3
Diagnostics Page
The service platform uses a Web-controlled user interface to access common service components and
perform diagnostics. There are two types of diagnostics:
•
Interactive:
The user is required to perform an operation on the ultrasound unit in order
for the test to be completed successfully. This option is not applicable when
used remotely.
•
Non-Interactive:
The tests are performed without the user’s intervention.
7-7-3-1
Diagnostics Page Overview
The left pane of the Diagnostics page is shown in
.
There are five main categories:
•
Calibration
•
Full System Test
•
Front End
•
Monitoring
•
BEP
You can expand each category to drill down to the various diagnostic tests by clicking on the plus sign
adjacent each category. For an example, see
.
When a diagnostic is selected in the left pane of the
Diagnostics
page, the corresponding tests in the
right pane can be performed (refer to the procedure instructions). For example, under Calibration, the
DC Offset Calibration diagnostic test can be performed (see
Figure 7-114 Diagnostics Page - Left Pane