Section 1-4 - Product Labels and Icons
Note:
When charging the battery for the first time after long-term storage, recover the battery
pack to original performance through repeating several cycles of full charging and
discharging.
NOTE:
When shipped, the Vivid i
n/ Vivid q N
battery is in a state of being approximately 30% charged. Before use, it is
necessary to fully charge and discharge the battery pack up to 3 times, in order to utilize Li-lon smart packs.
Important:
•
For Vivid
i
n and Vivid q N portable ultrasound scanners supplied from the factory with a dummy
battery, when unpacking the unit,
do not discard the dummy battery
.
•
When charging the Vivid
i
n or Vivid q N lithium ion battery on an external charger, since the battery
in fact forms one of the four legs of the scanner, removing it will leave the unit standing unbalanced.
For this reason, the dummy battery should be kept, so it may be inserted in position to provide
stability to the scanner while the lithium ion battery is being charged.
•
In addition, use of the dummy battery is recommended during transportation or long-term storage
of the Vivid
i
n and Vivid q N portable ultrasound scanner.