GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-83
8-4-12-4
Gas Spring Installation Procedure
NOTE:
When packed, the Gas Spring replacement part is fully extended. The spring must be
compressed prior to installation.
1.) Compress the Gas Spring by pressing downwards on the Gas Spring with its pressure release
button pressed against a solid surface until the Gas Spring is full compressed.
2) Turn the Gas Spring locking nut counter-clockwise until it rises about a quarter-way upwards on the
thread.
3) Insert the Gas Spring handle into the Gas Spring Housing Assembly until it is flush with the outer
cylindrical surface of the housing assembly.
4) Attach the Gas Spring to the housing assembly by turning it clockwise until the locking nut makes
contact with the housing assembly - see
Figure 8-95 Compressing the Gas Spring
Figure 8-96 Gas Spring Holder Inserted Flush with Housing Assembly
Pressure Release Button
Solid Surface
Press the Gas Spring downwards
in the direction of the arrow
Gas Spring Handle
Handle-end Flush with Housing Assembly
Lock Nut
Содержание Vivid in