Chapter 6 - Service Adjustments
6-5
6-2-5
Blue Tint "2" Adjustment Procedure
NOTE:
Systems with BT11 installed software have a Blue Tint "2" option that enables blue tint enhancement.
This option can be switched off if preferred by the user.
NOTE:
This option applies to the following LCD monitor installations:
System:
Vivid i n
Monitor:
BOE HYDIS-1
NOTE:
On older Vivid i/q systems where the IDTECH-1 monitor is supplied, the Blue Tint "2" option is not
applicable.
NOTE:
The Blue Tint "2" option is incompatible with IDTECH-1/2 monitors, therefore, on these systems, the
Blue Tint "2" option must be turned off.
To identify which monitor is installed on the system
:
1.) Log on as Administrator.
2.) Press Config > About > and select the
HwVersion
tab.
The monitor type appears under Module Name and the part number indicates the monitor type
- see
.
Figure 6-6 Identifying Monitor Type