Section 8-4 - SafeLock Cart Components Replacement
5) Align the lower Gas Spring connecting bracket with the SafeLock chassis until it is aligned with the
screw aperture on the chassis.
6) Insert the lower Gas Spring connecting screw and tighten it securely.
7) Replace the Main Harness cable by coiling it once around the Gas Spring and then reconnect it to
the AC unit.
8) Secure all cables in position with new cable ties.
9) Return the Lower Cover (Rear), as described in the
Lower Cover (Rear) Installation Procedure on
10) Return the Upper Cover (Rear), as described in the
Upper Cover (Rear) Installation Procedure on
11) Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Figure 8-97 Lower Gas Spring Securing Bracket
Lower Gas Spring securing
bracket aligned with
SafeLock Cart chassis