Chapter 8 - Replacement Procedures
8-105
9.
When prompted, type the following password:
geulsservice
and press
<
Enter
>.
Note: when typing the password, use lower-case letters.
A message is displayed.
The CD then begins the software installation process and a progress indicator
is displayed, as shown below:
This stage of the process takes approximately 5 mins.
10.
Press <CTRL+ALT+DELETE>.
The system re-boots.
11.
Follow the instructions on the screen, clicking
OK
(or Cancel) when prompted.
The machine performs a number of reboots before completing the installation.
As the installation proceeds, the system displays various installation screens.
The installation process will take approximately 15 minutes -
DO NOT
ATTEMPT TO STOP THE INSTALLATION BEFORE COMPLETION!
When done, the following PMC Firmware upgrade message is displayed:
12.
Move cursor over the displayed window and press the
SET button.
The following message is displayed
Table 8-6
Vivid
i
n/ Vivid q N Software Re-Installation (cont’d)
Step
Instruction
Expected Result