Section 7-4 - Performing Front End (FE) Diagnostics
7) When finished reviewing the first report, trackball to the
Exit
button and press
Set
to close the View
Report window.
8) Select the second report (
HPF600
); the report is highlighted in the
Reports
list, and the
corresponding results are shown in the Report window below, as shown in
9) When done, trackball to the
Exit
button and press
Set
to close the View Report window.
10) After reviewing the test results, proceed as follows:
•
Re-boot the system (this is obligatory).
•
Perform calibration, as described in
DC Offset Calibration on page 7 - 25
•
If the test passed but the original problem persists, contact your local On-Line Center for
assistance.
•
If errors were identified (results exceeding acceptable upper or lower limits), verify that all
system cables and boards are connected (not misplaced) and functioning properly, as
described in
- Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of that board is faulty. Continue trouble-shooting
down to the appropriate board. (If necessary, perform a specific test).
If the problem still persists, contact the GE HEALTHCARE On-Line Center for assistance
11.)Trackball to the
Exit
button and press
Set
to close the
View Test Log
dialog box.
12.)Press the <
Esc
> button on the keyboard to close the Diagnostic
Test Window
.
Figure 7-54 View Test Log - ACW HPF Test - 600