Chapter 8 - Replacement Procedures
8-137
11) Click
OK
.
The system should shutdown automatically.
(The screen will appear blank but fans will remain running and the On/Off key will be disabled).
12) Power-down the system by removing the AC adapter.
13.)Remove the Service Dongle.
14.)Wait approximately 5 seconds, then boot-up the system again.
Allow the system to boot up into
Normal Scanning
mode.
15) Press
Config
.
The system will prompt you to login.
16) Login as
ADM
.
The default password is
ulsadm
17) Press the
About
button and activate the
FwVersion
tab; verify that PMC version is now 23, as
shown in the example in
Figure 8-134 Checking the PMC Version
18) In the event that the PMC version is other than 23, burning has failed and it is necessary to repeat
above. Also, refer to the note below.