3-132
Section 3-10 - Connectivity Setup
3-10-8
eVue Dataflow Configuration
3-10-8-1
Introduction
It is possible to configure the Vivid
i
n/Vivid q N portable ultrasound scanner to allow the user to send
images or loops to a remote PC, in almost real time. When pressing Store, the loop will save not only
to the local archive (in raw-data format), but will also be transmitted to a designated PC (“e-Vue PC”)
on the network (in compact MPEG format), and will appear at the review PC within a very short time.
This feature proves very useful in the OR scenario, where consultation with another doctor in the
hospital
in real-time
is required.
3-10-8-2
Preparations
Before configuring the Vivid
i
n/ Vivid q N for eVue dataflow, it is necessary to decide upon a specific
PC in the hospital as being the e-Vue PC station. The service technician then configures the
Vivid
i
n/ Vivid q N to send images to that specific e-Vue PC, as described below.
3-10-8-3
Configuring the Vivid
i
n/ Vivid q N for eVue Dataflow
1.) Verify which PC in the hospital/institute has been designated as the e-Vue station.
2.) Make sure the Vivid
i
n/ Vivid q N is connected to the Network via cable or wireless.
3.) On the Vivid
i
n/ Vivid q N scanner, press the
Config
button.
The System Configuration window opens.
4.) Click
Connectivity
.
The Connectivity dialog box opens.
5.) Select the
Dataflow
Tab, as shown in