Chapter 8 - Replacement Procedures
8-53
8-3-8
LCD Display Replacement Procedure
8-3-8-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Display replacement
procedure.
8-3-8-2
Time Required
15 minutes
8-3-8-3
Preparation
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-3-8-4
LCD Display Removal Procedure
1) Remove the bearing handle, as described in the
Bearing Handle Removal Procedure on page 8 - 3
.
2) Remove the Control Panel and Keyboard as described in the
Control Panel and Keyboard Removal
3.) Remove the LCD Display Frame as described in the
LCD Display Frame Removal Procedure on
4.) Loosen and remove the four screws that fasten the LCD Display to the Rear Cover & Latch
Assembly, as shown in shown in
.
5.) Disconnect the LCD Flex ribbon cable from the BEP, as shown in
Use a flat screwdriver to carefully release the cable from the connector.
Figure 8-62 Removing the Screws from the LCD Display
Figure 8-63 Disconnecting the LCD Flex Ribbon Cable